CPACINC.COM

Computer and IT Solutions Professionals

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AUTORANGING INSULATION TESTER & RS-232 Window Version

MODEL TES-1602
  • 4,000 MW/V Reading/10,000MW analog indication
  • "Test Hold" function easy to catch up the value
  • Backlight illumination
  • Innovational RS-232 interface

CAT II pollution 2 1000V
CAT III pollution 2 600V

TES-1602
FEATURES¡G
¡¯ Via RS-232 interface to talk with PC (Window/DOS Versions)
¡¯ 3-3/4 Digital LCD, 4000 Count
¡¯ Auto Range for Mega Ohm range
¡¯ 50 segments Log Scaled Analog Display
¡¯ Easy to carry testing package
¡¯ Tests insulation at rated voltage into a 1mA load
¡¯ 200mA short circuit continuity current
¡¯ Meet VED0413

ELECTRICAL SPECIFICATIONS :
¡° AC Voltage
Range: 600V ACV
Resolution: 0.1Volt
Accuracy: 1.5%rdg+3dgt (1V¡ã600V)
¡° Continuity
Range: 40 Ohm (Ish>200mA )
Resolution: 0.01W
Accuracy: 1%rdg+30dgt
¡° Mega Ohm
Range:
4000MW/250V w/ autorange 4000M/400M/40M/4M
4000MW/500V w/ autorange 4000M/400M/40M/4M
4000MW/1000V w/ autorange 4000M/400M/40M/4M
Resolution:
4000MW/250V, 1count/1KW
4000MW/500V, 1count/1KW
4000MW/1000V, 1count/1KW
Accuracy: 3%rdg+5Count (under 1GW)
5%rdg+3Count (under 4GW)
GENERAL SPECIFICATIONS :
Display:
80 x 50mm Big LCD penal with 50 segments analog display
Max Reading 4000
Over range Indication :
"OL" will be shown on the LCD penal when out of range
measurement is made.
Low Battery Indication:
The will be shown when the battery need to be changed
Sampling Rate:
2.5 times/sec Digital Display
10 times/sec Bar Graph
Power Source:
Eight 1.5Volt AA size Battery
Operating Temperature and Humidity:
0oC to 40oC (32 oF to 104oF ), below 80% RH
Storage Temperature:
-10oC to 60oC ( 14 oF to 140oF )
Dimension:
200(L)x140(W)x95(H)mm
Weight:
Approx. 900g (with battery)
Accessories:
Test Lead, Battery, Instruction Manual